Flow Sensor: Ultrasound FlowSensor
Flow:
- Range
0 - 20 L/s
- Accuracy
<+/- 1%
- Resolution
1 ml/s
- Volume:
Range0
- 20 L
- Accuracy
+/- 5 m
- Resolution
l 1 ml
Measurement principle flow and density: ultrasound transit time measurement
Sampling rate: 2.000 Hz (1.000 measurement point)
Flow resistance at 1 l/s: practically negligible
Technical specification meets or exceeds ATS/ERS clinical performance standard
The SpiroScout® measures with a built-in ambient module automatically temperature, ambient pressure and relative humidity Power consumption: 2,5 Watt max
Dimensions complete: (T x B x H): 9 x 17,5 x 9 cm
Total weight: ca. 960 g
Material housing: ABS
Material disposable breathing tube: Polyethylene
PE Software: The SpiroScout® uses the well proven Ganshorn PFT Software LFX.
Minimum requirements for LFX:
Operating system Windows® 7, CPU minimum Intel® Core 2 Duo, 2 GB RAM, 40 GB HDD Alternative to LFX
Minimum requirements for LF8: