PROBE TONE
• 226 Hz, 85 dB SPL
• 1000 Hz, 75 dB SPL for Admittance (Y) curve with added Susceptance (B) and Conductance (G) curves
• The probe is supplied in handheld mode
INTENSITY
• 226 Hz: 85 dB SPL ± 2 dB
• 1000 Hz: 75 dB SPL ± 2 dB
• Frequency Accuracy: ± 0,5%
• Harmonic Distortion (THD): Less than 1%
ADMITTANCE MEASUREMENTS
• Compliance range at 226 Hz: 0.05 up to 7 ml
• Sensitivity scale: Auto Scales to appropriate range, available scale at 226 Hz 1, 2, 5 and 7 ml
• 1000 Hz: 0 to +25 mmho
• Sensitivity scale: Auto Scales to appropriate range, available scales 5, 10, 15, 20, 25 mmho
AIR PRESSURE
• Control: Automatic
• Range: two selections available: +100/-200 daPa or +200/-400 daPa
• Pressure accuracy: ±10 daPa or ±10%
• Sweep rate: automatic adaptive speed control or fast (200 daPa/sec)
• Indicator: Measured value is displayed
• Safety limitations: -750 up to +550 daPa
ACOUSTIC REFLEX TESTS
• Reflex test method: Threshold and Fixed level
• Stimulus duration: 0.5 or 1 sec.
• User selectable protocols for all test methods
• Ipsi or Contralateral stimulation for all reflex test
• Automated peak search functions available for Quick Test
FREQUENCIES AND INTENSITY RANGES
Ipsilateral:
• Pure tone level range (dB HL) from 50 to 100
• Frequency: 500, 1000, 2000, 4000 Hz
• Accuracy: ±1%
• Harmonic distortion (THD): less than 3%
Contralateral:
• Pure tone level range (dB HL) from 50 to 115
• Frequency: 500, 1000, 2000, 4000 Hz • Accuracy: ±1%
• Calibration accuracy ±3 dB
• Level steps: 5 or 10 dB
• On/Off ratio: 70 dB minimum